Advances in Diagnosis, Surveillance, and Monitoring of Zika Virus: An Update

  • 1ICAR-Indian Veterinary Research Institute, Bareilly, India
  • 2Division of Pathology, ICAR-Indian Veterinary Research Institute, Bareilly, India
  • 3Central University Laboratory, Tamil Nadu Veterinary and Animal Sciences University, Chennai, India
  • 4Department of Veterinary Microbiology and Immunology, College of Veterinary Sciences, UP Pandit Deen Dayal Upadhayay Pashu Chikitsa Vigyan Vishwavidyalay Evum Go-Anusandhan Sansthan, Mathura, India
  • 5Department of Biochemistry and Genetics, Barkatullah University, Bhopal, India
  • 6School of Engineering and Science, Tecnologico de Monterrey, Monterrey, Mexico
  • 7Division of Biological Standardization, ICAR-Indian Veterinary Research Institute, Bareilly, India
  • 8Laboratorios Lokímica, Departamento de Investigación y Desarrollo (I+D), Valencia, Spain

Zika virus (ZIKV) is associated with numerous human health-related disorders, including fetal microcephaly, neurological signs, and autoimmune disorders such as Guillain-Barré syndrome (GBS). Perceiving the ZIKA associated losses, in 2016, the World Health Organization (WHO) declared it as a global public health emergency. In consequence, an upsurge in the research on ZIKV was seen around the globe, with significant attainments over developing several effective diagnostics, drugs, therapies, and vaccines countering this life-threatening virus at an early step. State-of-art tools developed led the researchers to explore virus at the molecular level, and in-depth epidemiological investigations to understand the reason for increased pathogenicity and different clinical manifestations. These days, ZIKV infection is diagnosed based on clinical manifestations, along with serological and molecular detection tools. As, isolation of ZIKV is a tedious task; molecular assays such as reverse transcription-polymerase chain reaction (RT-PCR), real-time qRT-PCR, loop-mediated isothermal amplification (LAMP), lateral flow assays (LFAs), biosensors, nucleic acid sequence-based amplification (NASBA) tests, strand invasion-based amplification tests and immune assays like enzyme-linked immunosorbent assay (ELISA) are in-use to ascertain the ZIKV infection or Zika fever. Herein, this review highlights the recent advances in the diagnosis, surveillance, and monitoring of ZIKV. These new insights gained from the recent advances can aid in the rapid and definitive detection of this virus and/or Zika fever. The summarized information will aid the strategies to design and adopt effective prevention and control strategies to counter this viral pathogen of great public health concern.

Introduction

Zika virus (ZIKV) is the latest emergent virus after the Ebola epidemic. In the past, ZIKV has only been associated with mild disease; however, after subdue activity for six decades, it recently emerged as a significant threat to human health, with evident fetal abnormalities, microcephaly, serious neurological complications, and autoimmune disorders such as Guillain-Barré syndrome (GBS) (Al-Qahtani et al., 2016; Cao-Lormeau et al., 2016; Carteaux et al., 2016; Mlakar et al., 2016; Singh et al., 2016, 2017). ZIKV is a mosquito-borne virus in the genus Flavivirus of the Flaviviridae family that is transmitted by Aedes aegypti and Ae. albopictus mosquitoes, with the potential for rapid spread (Musso and Gubler, 2016; Weaver et al., 2016). ZIKV is an enveloped, single-stranded RNA virus that is closely related to dengue virus (DENV), yellow fever virus (YFV), and West Nile virus (WNV) (Fajardo et al., 2016; Kruger, 2016). The first isolation of ZIKV from the blood of a rhesus monkey from the Zika forest in Uganda was reported in 1947 (Hayes, 2009). There is increasing evidence of the placental transfer of ZIKV, as well as its ability to affect neuronal tissue of a growing fetus (Martines et al., 2016; Mlakar et al., 2016). Autoimmune complications like GBS and maternal transfer to fetuses resulting in microcephaly has become particularly alarming. A lack of suitable vaccines and drugs has complicated the response to the recent epidemic. The presently available treatment options remain supportive in nature. Hence, regular surveillance is advised, especially for travelers and pregnant women (Yadav et al., 2016). Furthermore, predisposing factors like climate change, globalization, an explosive population rise, and increased urbanization have aided the spread of this virus, posing pandemic potential (Haug et al., 2016; Troncoso, 2016).

Various platforms are available for ZIKV diagnosis. Isolation remains a difficult task, as it requires adequate biocontainment facilities to handle the samples. Hence, other assays like reverse transcription-polymerase chain reaction (RT-PCR) and enzyme-linked immunosorbent assay (ELISA) have been developed for ZIKV diagnosis (Landry and St. George, 2017; Rather et al., 2017). RT-PCR can be performed using urine, blood, and cerebrospinal fluid (CSF) from an early stage in infection. IgM and IgG ELISAs are available for diagnosis of ZIKV antibodies in serum (Shan et al., 2017a). Conventional diagnostics have the limitation of cross-reactivity, because ZIKV is genetically similar to other flaviviruses. Investigators have proposed several new diagnostic tools for early and accurate diagnosis of ZIKV infection (Miller et al., 2017; Sharma and Lal, 2017). Newer and faster techniques like biosensors, loop-mediated isothermal amplification (LAMP), and lateral flow assays (LFAs) have been developed recently, and they are being validated for commercialization (Sakudo et al., 2016). In addition, to spread awareness about ZIKV, people are sharing relevant photographs on popular social media platforms such as Pinterest (San Francisco, CA) and Instagram (Menlo Park, CA) as adjunct health educational tools (Fung et al., 2017). The objective of present review is to outline the recent technological progresses in designing and developing rapid and confirmatory diagnostic, surveillance, and monitoring approaches to counter ZIKV. This review brings up-to-date information on new achievements in the field of ZIKV diagnosis, and opines the core features of the in-use and forthcoming assays. The modern assays are speedy as well possess higher competence befitting the clinical diagnosis. Here, clinical to innovative means for rapid ZIKV detection and identification, precisely, clinical, gross and histopathological lesion-based diagnosis, and detection methods involving isolation of ZIKV, serological assays viz. multiplex microsphere immunoassay (MIA), Zika IgM antibody capture enzyme-linked immunosorbent assay (Zika MAC-ELISA), reporter virus neutralization test (RVNT), plaque reduction neutralization test and molecular assays like (RT)-PCR, real-time qRT-PCR, reverse transcription strand invasion based amplification (RT-SIBA) assay; nucleic acid sequence-based amplification (NASBA) coupled with CRISPR detection module, real-time loop-mediated isothermal amplification (RT-LAMP), transcription-mediated amplification (TMA) technology, reverse transcription isothermal recombinase polymerase amplification (RPA), flow cytometry, surface plasmon resonance-based technology, are reviewed and discussed.

Trends and Advances in Diagnosis, Surveillance, and Monitoring of ZIKV

The clinical signs of Zika fever are not pathognomic, therefore, confirmation of ZIKV infection requires isolation and identification of the virus. Diagnosis of ZIKV infection can be made based on apparent pathology, presence of the vector in the area, ZIKV-associated neurological syndromes, epidemiological evidence, and results of serological and molecular tests (Singh et al., 2016). ZIKV has been shown to be transmitted through body fluids during sexual intercourse and vertically, from mother to fetus (Balm et al., 2012; Gourinat et al., 2015; Leung et al., 2015; Musso et al., 2015; Tognarelli et al., 2015; de M. Campos et al., 2016; Jacob, 2016; Moulin et al., 2016; Shinohara et al., 2016; Staples et al., 2016; Pyzocha et al., 2017). Clinical samples that can be used for diagnostic purposes comprises of serum, the umbilical cord of an infant, urine, nasopharyngeal swabs, saliva, brain tissue, amniotic fluid, CSF, and placenta. In infected infants, ZIKV can be identified by immunohistochemical staining of the antigen in the umbilical cord and placenta (Staples et al., 2016). Diagnosis of ZIKV-infected stillborn fetuses can be made by histopathological and immunohistochemical staining (Landry and St. George, 2017). As mentioned earlier, most in-practice laboratory diagnostic tools include RT-PCR, multiplex PCR, antibody-capture ELISA, sandwich ELISA, and indirect immunofluorescent test (IIFT) (Steinhagen et al., 2016; Miller et al., 2017). Nucleic acid detection methods have been found most reliable, quick, sensitive, specific, and economical. Other molecular and serological tests can be used to confirm ZIKV infection (Staples et al., 2016). Differential diagnosis is an essential part of successful ZIKV detection as most of the time they are confused with Dengue virus at the early part of infection. ZIKV has attracted interest among researchers only in the recent years hence due to lack of laboratories that can accurately diagnose ZIKV most of the occasions they are not diagnosed properly (Hamel et al., 2016; Shankar et al., 2017).

Clinical, Gross, and Histopathological Lesion-Based Diagnosis

Viruria has been detected for more than 15 days after the onset of symptoms (Rozé et al., 2016). The clinical signs of ZIKV infection like fever, headache, myalgia, arthralgia, and maculopapular rash are inconsistent, as only one out of 4–5 people exhibit them. Hence, diagnosis based on clinical signs may not be reliable (Eppes et al., 2017). ZIKV replicates in virus-induced membranous replication factories (RFs). Endoplasmic reticulum of ZIKV infected human hepatic cells and neural progenitor cells show invaginations with pore-like openings toward the cytosol. Upon electron microscopy, it was revealed that the infection with ZIKV cause drastic change in microtubules and intermediate filaments organization forming cage like structure. Also, cytoskeleton-targeting drugs affect ZIKV infection severely and tight linking of ZIKV RF with host cell cytoskeleton changes is indicative of infection (Cortese et al., 2017). Leukopenia and thrombocytopenia are less common, whereas peripheral edema and conjunctivitis are more commonly encountered in ZIKV infections (Ioos et al., 2014; Rafiei et al., 2016). Occipitofrontal circumference can indicate microcephaly. To detect microcephaly in pregnant women, ultrasound testing is used (Lazear and Diamond, 2016; Oliveira Melo et al., 2016; Society for Maternal-Fetal Medicine (SMFM) Publications Committee, 2016). Epidemiological data is collected from maternal-fetal obstetrician/gynecologists, pregnant patients, and Zika-suspected cases to determine the threat of congenital ZIKV infection. The various kinds of deformities are based on the time of ZIKV exposure and also on the replication of ZIKV in the fetal and placental tissues. These data-gatherer techniques overcome the limitations of ultrasound-based strategies, where diagnosis depends exclusively on the detection of microcephaly and offer an alternative to detect malformations (Eppes et al., 2017).

Neuroimaging Diagnosis

Neuroimaging [computed tomography (CT) and magnetic resonance imaging (MRI) scans] can reveal congenital microcephaly (de Fatima Vasco Aragao et al., 2016). Abnormalities include atrophy of the brain parenchyma with secondary ventriculomegaly; calcification of the intracranium; cortical development malformations such as polymicrogyria and lissencephaly-pachygyria; hypoplasia of the corpus callosum, cerebellum, and brainstem; and hearing loss (Figure 1). Other abnormalities associated with ZIKV infection, including ocular abnormalities and arthrogryposis, can be found in infected fetuses. Post-natal (acquired) ZIKV infection can result in either a mild symptomatic or asymptomatic course, whereas prenatal (congenital) ZIKV infection can be severe, leading to brain abnormalities known as congenital Zika syndrome (Zare Mehrjardi et al., 2016, 2017). Intra-uterine ZIKV infection is characterized by reduced cortical gyration and white-matter hypomyelination or dysmyelination with cerebellar hypoplasia in majority of fetuses and newborns (Araujo Júnior et al., 2017). A marked ZIKV associated feature is abnormal head shape. Skull gains a collapsed shape with everted and/or cupped sutures with overriding bones in the occipital region leading to formation of redundant and folded skin. Calcified skull also needs to be taken into account for post-natal assessment (de Oliveira-Szejnfeld et al., 2017). A recent neuroimaging study of 17 fetuses with confirmed ZIKV showed that microcephaly occur in fetus at a minimum of 15 weeks time period requirement after a pregnant woman gets infected with ZIKV (Parra-Saavedra et al., 2017).

FIGURE 1
www.frontiersin.org

Figure 1. Diagnosis of microcephaly caused by Zika virus.

Isolation of ZIKV

ZIKV from a mosquito can be isolated in newborn Swiss albino mice via intraperitoneal, intracerebral, and subcutaneous inoculation (Marchette et al., 1969; Way et al., 1976). Hemagglutination inhibition test can be used to confirm virus in brain-passaged material (Haddow et al., 1964). For in vitro cultivation of the virus, Vero cells, rhesus monkey kidney cells (LLC- 192 MK2), and mosquito-origin [A. albopictus (C6/36), A. psuedoscutellaris (MOS61 or AP-61)] cell lines are employed (Barzon et al., 2016; Waggoner and Pinsky, 2016). Virus isolates from urine and saliva samples were recovered in Vero cell lines in one study, showing that Vero cell lines can be used to isolate ZIKV (Bonaldo et al., 2016). Another comparative study revealed that culture of ZIKV in C6/36 and Vero cells yield higher virus titers (Coelho et al., 2017). BHK21 suspension cell lines were also studied for the propagation of ZIKV to aid in ZIKV vaccine production (Nikolay et al., in press). Because live ZIKV sample handling requires specialized facilities to prevent its spread to handlers, not all laboratories can isolate the virus. Low-level viremias can also limit the opportunity for isolation of the virus (Lanciotti et al., 2008; Zanluca and Dos Santos, 2016). To test vaccines, several laboratory animal models of ZIKV infection have been evaluated. Two mouse models were studied, namely A129 type I interferon receptor knockout mice and AG129 type I and type II interferon receptor knockout mice. Three-week-old A129 and AG129 mice showed neurological signs and the presence of virus in their brains 7 days after challenge, whereas immunocompetent mice did not produce similar signs when challenged with ZIKV (Brault and Bowen, 2016). In one study, 15 non-human and 18 human cell lines were used to determine the replication of ZIKV in various cells. Results revealed that ZIKV replicated in neuronal (SF268), retinal (ARPE19), pulmonary (Hep-2 and HFL), hepatic (Huh-7), placental (JEG-3), muscle (RD), and colonic (Caco-2) cell lines. Replication in placental cell lines shows the ability of ZIKV to cross the placental barrier. ZIKV replication was also found in non-human cell lines, namely pig (PK-15), chicken (DF-1), non-human primate (Vero and LLC-MK2), rabbit (RK-13), and hamster (BHK21) cell lines; hence, these animals have been suggested as animal models for ZIKV (Chan J. F.-W et al., 2016).

Serological Tests

Hemagglutination inhibition, serum neutralization, and complement fixation are the useful tests for the diagnosis of Zika fever (Fagbami, 1979; Monath et al., 1980). Short viremias make it difficult to detect ZIKV; hence, serological diagnosis is a better way out to determine the status of an individual for a longer period of time (Shan et al., 2017a). Mouse protection test has also been employed to identify ZIKV antibodies in serum samples (Kirya and Okia, 1977). Monitoring of IgG and IgM antibodies in serum by ELISA is highly useful for serological diagnosis of ZIKV (Staples et al., 2016). Seroconversion is confirmed by analyzing the antibodies titers in paired serum samples (acute and convalescent phase) during the course of infection. Recently neutralization assay was combined with real time PCR platform to detect the end point so that there was 100% sensitivity and specificity of detection of ZIKV. Authors suggest this assay can be a confirmatory test for the samples positive by commercial immunoassays for detection ZIKV antibodies (Wilson et al., 2017).

Plaque Reduction Neutralization Test

Plaque reduction neutralization test (PRNT) is used to detect antibodies against ZIKV and shows less cross-reactivity with antibodies against other Flavivirus members than ELISA (Petersen et al., 2016; Staples et al., 2016). PRNT is a labor-intensive technique requiring a week for interpretation of results. Hence, a reporter virus possessing the luciferase gene from Renilla was used in a neutralization assay, and it showed similar specificity as PRNT, but reduced the time of diagnosis to 2 days (Shan et al., 2017a).

Reporter Virus Neutralization Test (RVNT)

RVNT uses the same principle as that of PRNT but the difference in this test is the use of luciferase tagged ZIKV thus allowing the quantification of neutralization within 24 h instead of 7 days in PRNT. Similarly, this assay can be done on 96 well plates hence this assay minimizes the testing time and is a high throughput assay (Shan et al., 2017b).

Detection of Specific IgG and IgM Antibodies

Anti-ZIKV IgM can be detected after 5 days of onset of clinical symptoms and can last up to 3 months (Calvet et al., 2016). A recombinant ZIKV non-structural protein 1 (NS1) based-ELISA has been developed as an additional serological diagnostic which shows very low cross-reactivity with DENV antibodies, along with high specificity for ZIKV, and can be applied for serum-based diagnosis in pregnant women, travelers of ZIKA-endemic regions and individuals suspected of infection for counseling. The sensitivity was 58.8, 88.2, and 100% for IgM, IgG, and IgM/IgG respectively with 99.8% specificity (Steinhagen et al., 2016). Detection of specific IgG antibodies in maternal serum by ELISA represents an important approach to diagnose ZIKV infection and helps in determining the mechanism of neonatal microcephaly in congenital infection, even in mothers where ZIKV RNA has cleared (Sumita et al., 2016). A study using this ELISA was done on 105 serum samples collected from travelers returning to Israel. The sensitivity of the IgM ELISA differed among patients of Israeli origin and from European origin; hence, to minimize false negatives, further diagnostic tests are needed (Lustig et al., 2017). ZIKV-specific epitope-based serological assays can prevent the problem of cross-reactivity with other flavivirus antibodies (Landry and St. George, 2017). The IgM ELISA is a U.S. Food and Drug Administration (USFDA) approved assay for use with serum and CSF (Murray et al., 2017). A comparison of the commercially available NS1-based ELISA and the IgM ELISA showed better specificity with the IgM ELISA, but sensitivity was suboptimal during the first 5 days of infection (Kadkhoda et al., 2017). The competitive ELISA developed against NS1 of ZIKV, where out of confirmed RT-PCR positive 158 sera/plasma samples, 91.8% samples yielded higher than 50% inhibition (Balmaseda et al., 2017). The assay is robust, low cost strategy to surveillance program and seroprevalence studies and quickly adapted in ZIKV endemic areas.

Zika IgM Antibody Capture Enzyme-Linked Immunosorbent Assay (Zika MAC-ELISA)

The FDA has approved the use of CDC Zika IgM Antibody Capture Enzyme-Linked Immunosorbent Assay (Zika MAC-ELISA) for antibody testing in case of emergency. The technique is quantitative detection of ZIKV IgM antibodies in serum or CSF collected from patients. Result interpretation in MAC ELISA is based on the positive/Negative ratio (P/N) and the samples greater than or equal to 3 are considered to be “presumptive Zika IgM positive.” These samples are sent to approved laboratory for further confirmation by PRNT (Cordeiro, 2017; Landry and St. George, 2017). False positive results may be ruled out by checking with other methods like PRNT against Zika, dengue, and other flaviviruses (FDA, 2016).

Multiplex Microsphere Immunoassay (MIA)

The MIA detects antibodies against ZIKV surface protein E, non-structural NS1 and NS5 proteins in order to increase the accuracy and speed. It possesses improved serological diagnostic capability for ZIKV. MIA allows incorporation of more than one antigen to enhance the diagnostic coverage. The diagnostic width of MIA may further be improved by adding multiple antigens from ZIKV, DENV, YFV and other flaviviruses circulating in the same geographical area (Wong et al., 2017). A MIA has been developed for detection of 6 flaviviruses, 6 alphaviruses, and 1 bunyavirus of human importance (Basile et al., 2013). In this multiplex assay, ZIKV antigens may also be incorporated for developing ZIKV diagnostics.

Miscellaneous

A recently developed experimental murine model and ZIKV infection model in cynomolgus macaques may help to elucidate disease course, as well as would be useful in testing the anti-ZIKV drugs and vaccines (Koide et al., 2016; Rossi et al., 2016). In addition, a 3-[4,5-dimethyl-2-thiazolyl]-2,5-diphenyl-2H-tetrazolium bromide-(MTT)-based cell viability assay was developed to assess cell death in human and monkey cells caused by ZIKV. This has been claimed to be economic, simple, and rapid colorimetric assay that could be used for quantification of neutralizing antibodies against ZIKV (Müller et al., 2017).

The U.S. Centers for Disease Control and Prevention (CDC) Zika MAC-ELISA, InBios ZIKV Detect™MAC-ELISA, and Euroimmun anti-ZIKV IgM ELISA were compared in a study, and the CDC and InBios MAC-ELISAs were found to exhibit comparable sensitivities, whereas the Euroimmun ZIKV IgM ELISA was less sensitive (Granger et al., 2017). Recently, Euroimmun IgG/IgM ELISA was compared with PRNT and MAC-ELISA where it was found that Euroimmun ELISA had 92.5% specificity and 39.5% sensitivity when compared with MAC ELISA, while specificity ranged from 65 to 81% and sensitivity 83–92% when compared with PRNT (L'Huillier et al., 2017a); whereas Pasquier et al. (2018) revealed Diapro ZIKV IgG or IgM ELISA immunoassays more sensitive than the Euroimmun immunoassays.

Researchers have reported that MIA is more sensitive than IgM-capture ELISA for serological diagnosis of ZIKV infection. Moreover, the easy preparation, smaller sample volume requirement, and short time to get results are significant advantages of the MIA. Therefore, the proposed MIA is a useful tool for assessing immune responses in vaccination trials and in clinical disease (Wong et al., 2017). Recently, a multiplex diagnostic assay to detect DENV and ZIKV IgG, IgA, and IgM antibodies on a plasmonic gold platform have been developed (Zhang et al., 2017). An IgG/IgM based LFA, the DPP Zika IgM/IgG assay, marketed by Chembio Diagnostic Systems, is also available in the market (Nicolini et al., 2017). Electro-generated chemiluminescence was linked with polystyrene beads, which were also conjugated with monoclonal antibodies (mAbs) against ZIKV. This conjugated structure was used to capture ZIKV from fluids and detect up to one plaque-forming unit (PFU) of ZIKV in 100 μL of fluid (Acharya et al., 2016). A pair of mAbs has been selected to construct a rapid diagnostic test in form of strip. The diagnostic strip was able to detect NS1 antigen is serum samples from various geographic areas in the Americas and India. The same test is linked with mobile phone camera and the images taken can be analyzed with ImageJ software, which allow objective analysis and eliminates user subjectivity in reading test results (Bosch et al., 2017).

Molecular Diagnosis

RT-PCR

The ZIKV RNA has been reported in various body fluids such as blood, saliva, urine, and amniotic fluid (Hills et al., 2016; Rather et al., 2017) and hence it can be detected by RT-PCR. In a comparison of RT-PCR testing for ZIKV RNA in urine, serum, and saliva specimens obtained from people traveling in Florida and Brazil, urine samples were found to be the most suitable specimen for identifying acute infections with ZIKV (Bingham et al., 2016; Lamb et al., 2016). The RealStar Zika RT-PCR Test Kit, marketed by Altona Diagnostics, was found to be 91% sensitive and 97% specific in detecting ZIKV RNA; hence, this assay kit was proposed to diagnose early-stage ZIKV infection (L'Huillier et al., 2017b).

In a study, which tested a real-time RT-PCR based on the ZIKV envelope and NS2B genes was on urine and serum specimens from pregnant women, newborns, suspected symptomatic GBS patients, and travelers to ZIKV epidemic areas in New York. Due to very short viremia time period during ZIKV disease, there exists a challenge in selecting the appropriate sample for diagnosis, as ZIKV RNA may not be detected in serum or plasma collected 10 days after the onset of disease. Urine samples were found to be more reliable than serum samples, as the viral load was higher, and results of this assay were more dependable than those of serology (St. George et al., 2017). A study in Italy examined the whole blood, plasma and urine as samples for detection of RNA from 3rd day to 28 days post-ZIKV infection in 10 patients. Results showed that urine samples from all the individuals were ZIKV positive up to 21st day post-clinical sign while ZIKV could be detected in the whole blood up to 26 days (Rossini et al., 2017). During early disease, viral genomes could be detected in serum (Haug et al., 2016). ZIKV RNA has been detected in the breast milk of women, but there has been no report of transmission of ZIKV through breast milk. Another report showed that ZIKV persists for 81 days in whole blood, but 73 days in serum. Hence, whole blood can be used for ZIKV RNA detection in asymptomatic patients (Eppes et al., 2017; Murray et al., 2017). A dual target RT-PCR assay run on an automated m2000 system was developed to extract RNA from whole blood, serum, plasma, and urine and detect ZIKV by PCR. The detection limit of this assay was shown to be 120 copies/mL with whole blood, 30 copies/mL with serum, and 40 copies/mL with plasma and urine samples (Frankel et al., 2017).

Symptomatic pregnant women and affected persons display low-grade fever, maculopapular rashes, arthralgia, and non-purulent conjunctivitis for 2–7 days. Recently, in 2016–2017, asymptomatic pregnant women in American Samoa were tested under a syndromic surveillance program, with results displayed in electronic health records (Hancock et al., 2017).

A sensitive and specific one-step RT-PCR assay which includes GAPDH as internal control, which allows detection of PCR inhibitors in the sample, and rule out the possibility of false negatives was developed by Balm et al. (2012). The lower limit of detection of this assay was 140 RNA copies/PCR, with no cross-reactivity with closely related flaviviruses including DENV, YFV, and JEV.

A study was conducted using urine spiked with a commercially available heat-inactivated ZIKV panel (Exact Diagnostics® ZIKV Verification Panel), and several parameters were investigated, including the stability of ZIKV at different storage temperatures and the use of stabilizers for ZIKV RNA detection. The results showed that RNA degraded when stored at room temperature, whereas stability was better at 4°C. Storage of urine samples at −80°C led to decrease in RNA concentration. Similarly, the addition of nucleic acid stabilizers increased RNA recovery (Tan et al., 2017).

Real-Time RT-PCR

Real time PCR offers several advantages over conventional RT-PCR including rapidity, low false positive, higher sensitivity and specificity with quantitative analysis. Blood and urine samples were collected and tested by real-time RT-PCR to confirm ZIKV infection (Hancock et al., 2017). Real-time RT-PCR can detect ZIKV RNA at an early stage in infection (Faye et al., 2008, 2013). ZIKV RNA can easily be detected in urine samples using real-time RT-PCR, with the added advantage of easy, non-invasive sample collection. This is especially useful during an epidemic and for the testing of travelers (Gourinat et al., 2015). Based on real-time RT-PCR analyses, several researchers have documented that ZIKV loads are higher in plasma than in urine samples (Pessôa et al., 2016). The probes developed against NS5 region in a real-time PCR and the sensitivity was 32 genome-equivalents and 0.05 plaque forming unit (pfu). The assay was able to detect at least 37 ZIKV isolates representatives of wide geographic area in Africa and Asia over the period of last 36 years. Also, it could differentiate ZIKV among 31 other flaviviruses. A SYBR Green one-step real-time RT-PCR assay detected ZIKV RNA, even at a titer of 1 PFU/ml; therefore, it is very useful for diagnosis and surveillance of ZIKV infection (Xu et al., 2016). Another study compared nine real-time RT-PCR assays and found that some assays were not suitable for the specific detection of ZIKV (Corman et al., 2016). A quantitative RT-PCR targeting E gene has been very recently developed that could detect all Asian and African lineages of ZIKV. This assay alleviates the disadvantage of earlier developed real-time PCR as they could not detect all strains of ZIKV and, moreover, this assay also showed a detection limit of 5 RNA transcript copies (Yang et al., 2017). A list of ZIKV detection assays is presented in Table 1.

TABLE 1
www.frontiersin.org

Table 1. List of commercially available methodologies/kits for ZIKV diagnosis.

Real-time RT-PCR is a quick, reliable, sensitive and specific method. A comparison of 7 published real-time RT–PCR assays to determine the analytical sensitivity revealed presence of up to 10 potential mismatches in oligos with the Asian lineage. The two new assays developed by Corman et al. (2016) had 0–4 oligo mismatches. The continuous evolution in the viral genome and mismatch between oligos and genome may lead to reduced sensitivity. The study of 174 ZIKV genomes, exhibited the presence of mutation in diagnostic regions, however these were present in lower frequencies (Metsky et al., 2017). Hence, the diagnostic assay should be carefully chosen and need to be frequently evaluated and updated.

The co-circulation of other arboviruses hinders clear differential diagnosis of ZIKV (Pinto et al., 2015). A novel, highly sensitive and specific real-time RT-PCR assay that did not show cross-reactivity with other flaviviruses was developed by targeting the conserved 5′-untranslated region (5′-UTR), envelope (E), non-structural protein 2A (NS2A), NS5, and 3′-UTR of the ZIKV genome for improved laboratory diagnosis. It can detect even 5–10 RNA copies/reaction. Compared to the ZIKV-E gene targeting real-time PCR, the ZIKV-5'-UTR assay showed higher sensitivity in detecting ZIKV in most human tissues and the highest sensitivity in samples obtained from the testis/epididymis and kidney. This RT-PCR assay was found highly specific with no cross-reactivity with chikungunya virus, DENV, YFV, JEV, WNV, and hepatitis C virus (Chan et al., 2017).

Real-Time PCR Based Neutralization Assay

It is an assay into which the serum neutralization test and Real-time PCR tests are combined. The neutralization endpoint is measured by real -time PCR instead of counting plaques. The test takes 72 h to complete. This test may be employed as confirmatory test for those serum samples that are positive in an IgM/IgG ELISA. The sensitivity of this Real-time PCR based neutralization assay is 100% for both ZIKV and DENV (Wilson et al., 2017).

Surface Plasmon Resonance

Surface plasmon resonance-based technology for the sensitive detection of ZIKV RNA has been developed very recently wherein results revealed that bimetallic nanoparticle quantum dot-mediated fluorescent signals were stronger than those of single metal nanoparticles for the detection of ZIKV RNA (Adegoke et al., 2017). The technique is a reliable detection platform for rapid and ultra-sensitive detection of ZIKV genome. However, the higher cost of the test reduces its utility in low and middle-income group countries. A list of commercially available kits for ZIKV diagnosis is shown in Table 2.

TABLE 2
www.frontiersin.org

Table 2. List of commercially available kits for ZIKV diagnosis.

Flow Cytometry

Flow cytometry technique is also used for the detection of ZIKV NS3 antigen in the whole blood samples using polyclonal antibodies. It has been reported recently that most patients revealed a decrease in ZIKV antigen during later phases of the disease while some showed higher antigen level. Flow cytometric study also revealed that CD14+ monocytes are the targets of ZIKV during acute phase of infection (Lum et al., 2017). Immature dendritic cells present in human skin cells are permissive to ZIKV infection and 24 h post-infection express viral envelope when infected with 0.5 multiplicity of infection (MOI). The intracellular presence of the viral envelope protein is detected by using a broadly neutralizing Ab 4G2 by flow cytometry. Also, the transcription level of RIG-I, MDA5, IRF7 and TLR3 expression is upregulated as soon as 6 hpi which lasts for 48 h. The role of these genes as sensor of infection may also have a role in the detection of ZIKV (Hamel et al., 2015). In ZIKV-infected mice, CD4+ T cells which have encountered ZIKV antigens, exhibit a typical Th1 cytokine profile distinguished with the production of IFN-γ, TNF-α, and IL-2. In infected mice, a higher proportion of CD11a+CD49d+ CD4+ T cells express the Th1 transcription factor (T-bet) in comparison to mock-infected mice (Pardy et al., 2017).

DNA Sequencing

DNA sequencing provides confirmation of ZIKV infection (Zanluca et al., 2015). Sequencing of the NS5, NS3, and E genes can indicate the relationship between ZIKV strains (Fonseca et al., 2014; Grard et al., 2014; Tognarelli et al., 2015). Viral genomes may be enriched by multiplex PCR from samples containing as low as 50 copies of viral genome. Oligos were designed using Primal Scheme software and sequencing was done with the Oxford Nanopore MinION sequencing device. Clinical sample itself may be the starting material and within 1–2 days consensus viral sequence is obtained to do studies regarding evolution and spread of the virus (Quick et al., 2017). Multiple sequencing of 110 ZIKV genomes isolated from clinical and mosquito samples from 10 countries and territories, was done and compared with other 64 published genomes from NCBI GenBank which revealed that ZIKV outbreak has started from Brazil (Metsky et al., 2017). Deep sequencing can be employed to analyze the ZIKV genome, and a comparison with sequences available in a data bank can help to identify nucleotide changes in specific strains of the virus (Buechler et al., 2017). More recently, by employing next generation sequencing the whole genome sequence of ZIKV strain AFMC-U has been amplified from the urine sample of a Korean traveler. This report also shows the suitability and importance of urine as a sample for ZIKV diagnosis (Gu et al., 2017).

Other Advanced Diagnostics

Reverse Transcription Isothermal Recombinase Polymerase Amplification

A reverse transcription (RT) isothermal recombinase polymerase amplification assay (RT-RPA) that targets the NS2A region with 92% sensitivity and 100% specificity compared to RT-PCR has been developed (Abd El Wahed et al., 2017). It is a rapid assay which takes only 3–15 min and able to detect 21 RNA molecules. RPA shows many advantages over both real-time PCR and other isothermal amplification methods including Nucleic acid sequence-based amplification (NASBA), LAMP, Strand-displacement amplification (SDA), Rolling circle amplification (RCA), and Helicase-dependent amplification (HAD) in power saving (runs at 37°C), simple primer designing, quickness, no initial heating step and robustness for biological substances (Zaghloul and El-shahat, 2014).

Transcription-Mediated Amplification (TMA) Technology

The technology uses two enzymes, reverse transcriptase (RT) and RNA polymerase. It produces RNA amplicons, opposite to other assays which produce DNA and have rapid kinetics, which result in higher amplitude amplification within 15–60 min. The TMA technology-based Aptima Zika Virus assay (Hologic, Marlborough, MA) runs on automated Panther system and has shown 94.7% sensitivity and 94.8% specificity relative to that of real-time RT-PCR (Ren et al., 2017). A paper-based point-of-care test was developed very recently to detect multiple flaviviruses, and a microfluidic paper-based analytical device (μPAD) has been used to detect as little as 10 ng/ml of NS1 protein in plasma and blood. This assay is economic, specific, and provides a result in less than 8 min (Bedin et al., 2017).

Real-Time Loop-Mediated Isothermal Amplification (RT-LAMP)

Researchers have developed a one-step RT-LAMP technique that targets the entire ZIKV genome and shows high sensitivity and specificity. The RT-LAMP assay was compared with conventional and quantitative real-time RT-PCR, and results revealed excellent sensitivity and specificity in the detection of ZIKV (Wang et al., 2016). A LAMP assay to distinguish between African and Asian lineages of ZIKV has been developed to detect ZIKV RNA from mosquitoes, virus-spiked samples, human saliva, blood, plasma, and urine without RNA extraction (Chotiwan et al., 2017). The LAMP assay has the disadvantage of product carry-over contamination; hence, closed-tube techniques are usually preferred (Karthik et al., 2014). Recently, a LAMP assay to detect ZIKV was developed that utilizes a dTTP-dUTP mix with a thermolabile uracil DNA glycosylase that can prevent carryover contamination (Yaren et al., 2017). An RT-LAMP without a microfluidic cassette for the detection of ZIKV has been introduced as a point-of-care diagnostic assay (Song et al., 2016). An RT-LAMP assay combined with an LFA was developed to detect even a single copy of ZIKV RNA, with highly specific results available within 35 min (Lee et al., 2016). In addition, a smartphone-based RT-LAMP assay for easy detection of ZIKV in human saliva, urine, and blood was created. This assay used a portable LAMP box powered by a 5 V power source and a smartphone built using a chromaticity algorithm to scan for fluorescent signals. Use of this technology increased the sensitivity of this assay 5-folds compared to that of naked eye detection of colorimetric results (Priye et al., 2017). Finally, a portable device was developed that can extract nucleic acids using magnetic particles, then perform real-time RT-RPA or RT-PCR and interpret results by fluorescence detection. A 3D printer has also been used to process 8–12 samples for the detection of ZIKV, with fluorescence detection of amplified products by smart phone. This assay is cheap and portable (Chan K. et al., 2016). A higher sensitivity was obtained by Tian et al. (2016); they achieved a detection limit of one attomole using a synthetic ZIKV oligonucleotide within 27 min in serum.

Nucleic Acid Sequence-Based Amplification (NASBA)

Recently, a novel diagnostic assay has been developed that employs biomolecular sensors to colorimetrically detect ZIKV RNA by nucleic acid sequence-based amplification (NASBA). It also uses CRISPR technology that can differentiate between ZIKV strains. The coupled NASBA and CRISPR detection module was able to discriminate viral strains differing by one base and also able to differentiate the African ZIKV strain with that of American strains. The programmability of molecular sensors will help in addressing rapidly changing diagnostic requirements (Pardee et al., 2016).

Reverse Transcription Strand Invasion Based Amplification (RT-SIBA) Assay

RT strand invasion-based amplification by a battery-operated portable device for the diagnosis of ZIKV was reported to work well, even with incompletely purified RNA. It relies on a recombinase coated single-stranded invasion oligonucleotide purposed to separate complementary target duplex. Resulting single-stranded target template is extended by DNA polymerase at a relative low and constant temperature. The method is simple and can run on low cost equipments. The advantage of RT-SIBA includes no production of detectable fluorescence signal if the target template is absent, where in real-time RT-PCR a detectable signal is produced after 30 threshold cycles (Ct). Also, it is able to reproducibly detect as low as 10 copies of RNA (Eboigbodin et al., 2016).

Bioplasmonic Paper-Based Device (BPD)

The technique is based on the detection of IgG and IgM antibodies to NS1 protein of ZIKV. ZIKV-NS1 protein here acts as capturing element and gold nanorods act as plasmonic nano-transducers (Jiang et al., 2017). The technique offers and excellent stability of the BPD at room temperature and even at higher temperatures using metal–organic framework dependent biopreservation. The technique eliminates the storage and transportation of device at low temperature (Morrissey et al., 2017), thus having field applicability. The same device may be adapted for diagnosis of other infectious agents also.

Other Technologies

Other recent technologies for ZIKV diagnosis include paper disk tools, which work on the principle of a color change from yellow to purple in the presence of ZIKV antigen (Shukla et al., 2016). Recent techniques like “toehold switches” that can bind to any sense RNA sequences can also be applied for ZIKV diagnosis (Pardee et al., 2016). To concentrate ZIKV in samples with low viral loads, ultracentrifugation and polyethylene glycol precipitation can be performed, but these methods have their own limitations and can interfere with PCR assays (Novotny et al., 1992). Therefore, recently, magnetic nanoparticle-based concentration has been preferred. Nanoparticles like such as iron, nickel, and cobalt can be used, but their stability is poor; hence, they are encapsulated with silica, graphite, or a polymer (Saraswati et al., 2012). This method has been used for DENV detection using a PCR assay that can also be employed to detect ZIKV (Sakudo et al., 2016). A liposome-based biosensor that is cheap, portable, specific, and sensitive has already been developed for DENV (Zaytseva et al., 2004). These platforms can also be utilized for the detection of ZIKV. Very recently surface-enhanced Raman spectroscopy (SERS) was utilized to detect ZIKV and DENV in a multiplexed assay platform. To improve the sensitivity of LFA, SERS has been combined so as to increase the ZIKV diagnosis along with differentiation from DENV (Sánchez-Purrà et al., 2017).

Different diagnostic platforms available for Zika virus detection are presented in Figure 2.

FIGURE 2
www.frontiersin.org

Figure 2. An overview on different diagnostic platforms available for Zika virus detection.

ZIKV Surveillance Strategies

Extensive ZIKV epidemiological studies should be carried out to support advanced disease surveillance and monitoring approaches, and geographical information system (GIS) and appropriate networking programs should be employed to track the spread of the virus (Dhama et al., 2013; Shukla et al., 2016; Sharma and Lal, 2017). Based on the surveillance data of the Zika epidemics in the Tolima department, Colombia, epidemiological mapping has been developed employing GIS. Kosmo Desktop 3.0RC1® was used as the GIS software in this study and it was found that eastern area of this department was highly affected with ZIKV (Rodriguez-Morales et al., 2016). Similarly other reports also show that GIS was employed to develop epidemiological mapping in Valle del Cauca and Pereira department of Colombia. It was found that North part of Valle del Cauca and South-West part of Pereira department had most of the ZIKV burden (Rodriguez-Morales et al., 2017a,b). A mobile App (ZIKATracker, zikatracker.net) has recently been developed to report the occurrence of cases, which can be helpful for effective and timely control of the disease (Kelvin et al., 2016). Early reporting of ZIKV cases can allow the pattern of disease spread to be determined, as well as limit its further spread (Nishiura et al., 2016). Efficient serological and molecular detection techniques are warranted for improved surveillance to prevent the spread of disease by adopting timely and appropriate control measures (Waggoner and Pinsky, 2016; Sharma and Lal, 2017). In North America, the GeoSentinel Surveillance Network data platform has been used to track Canadian travelers displaying signs of acute ZIKV infection and fetal neurologic anomalies between October 2015 and September 2016 on the CanTravNet site (Boggild et al., 2017).

When the recent ZIKV outbreak occurred in South and Central American states, several national and international organizations expressed their concern for the safety of athletes, players, coaches, and viewers during the Rio Olympic and Paralympic games. Even blood and urine samples from asymptomatic persons were tested by real-time RT-PCR (Shadgan et al., 2016). Epidemiological studies confirmed that, after 2015, approximately 1.5 million people in Brazil alone were infected with ZIKV, with almost 80% of cases asymptomatic but positive for viral RNA. Since 2007, more than 55 countries in the Americas, Asia, Africa, the Caribbean, and the Pacific region have reported cases with GBS syndrome, retinal/eye lesions, and/or increased incidence of microcephaly that were confirmed by RT-PCR, IgM ELISA, or neutralizing antibody-based serological tests (Carod-Artal, 2016; Pyzocha et al., 2017).

Past vaccination against another flavivirus or a recent infection with a flavivirus may interfere with serological testing, increasing the probability of false positive results. In addition, these tests are time-consuming, laborious, and cross-reactivity with antibodies against other flaviviruses may lead to a misinterpretation of results. Hence, molecular tests are preferred over serological approaches. A number of nucleic acid amplification-based tests have been used to detect even acute infections by specifically replicating ZIKV RNA. However, in the case of asymptomatic patients or convalescent-phase samples, nucleic acid based techniques have limitations; therefore, specific and sensitive antibody-based tests that can detect ZIKV-specific epitopes or anti-ZIKV antibody are needed for diagnosis (Landry and St. George, 2017).

Advanced detection techniques include multiplex PCR, LAMP, recombinant diagnostics, biosensors/biochips, microarrays, and nanodiagnostics (Ratcliff et al., 2007; Bergquist, 2011; Kawadkar et al., 2011; Dhama et al., 2014; Van den Hurk and Evoy, 2015; Lambe et al., 2016). WHO declared the Zika epidemic a public health emergency of international concern, which emphasized need for rapid, accurate, cost-effective, and sensitive point-of-care diagnostics to prevent infection and its health effects of GBS and microencephaly in developing fetuses (Calvet et al., 2016; Shukla et al., 2016). Along with developing rapid and confirmatory diagnostics, discovering drugs, medicines, therapeutics, and vaccines, as well as effective prevention and control interventions, is necessary to safeguard the health of humans from this important pathogen of global importance (Koff et al., 2013; Singh et al., 2015; Khandia et al., 2017; Munjal et al., 2017a,b).

Conclusions and Future Perspectives

ZIKV existence is known since 1947, but, until recently, it has only been associated with mild disease. Nonetheless, its recent emergence has presented a pandemic threat world over with the appearance of increased cases of neurological anomalies like GBS and microcephaly. Hence, researchers across the world emphasized targeted to develop rapid confirmatory diagnostics and efficient drugs and medicines, as well as design appropriate prevention and control strategies to curtail ZIKV spread and associated ill effects. Understanding the factors responsible for the apparent greater pathogenicity of ZIKV infection by in-depth molecular exploration on the virus and host-pathogen interactions would help in designing effective diagnostics, drugs, and control strategies to counter ZIKV. Pre-existing antibodies to DENV may cross-react with ZIKV and, through the phenomenon of antibody-dependent enhancement, and result in an increase in the virus titer and enhanced ZIKV infection. Exploiting recent knowledge gained in immunology, biotechnology, and molecular biology, there is need to develop accurate and rapid detection assays for ZIKV/Zika fever, so as detailed disease informatics could be ascertained with an early and timely implementation of disease prevention and control measures. A better understanding of ZIKV pathogenesis and genetics would allow novel targets to be identified for the design and development of effective and safer drugs, medicines, pharmaceuticals, and vaccines to counter ZIKV effectively. Stronger and more extensive ZIKV surveillance, monitoring, and networking programs need to be implemented to prevent additional ZIKV emergencies in the future.

Author Contributions

All the authors substantially contributed to the conception, design, analysis and interpretation of data, checking and approving final version of manuscript, and agree to be accountable for its contents. RS designed an overview on virological and molecular diagnosis; KD, RT, KK, RK, and AM updated different aspects regarding serological and recent diagnostic advances; KK and RK designed tables and KK designed the figures; RS, KD, YM, HI, and RB-M finally reviewed, analyzed, edited and concluded the review compilation.

Funding

This compilation is a review article written, analyzed and designed by its authors and required no substantial funding to be stated.

Conflict of Interest Statement

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Acknowledgments

All the authors acknowledge and thank their respective Institutes and Universities.

References

Abd El Wahed, A., Sanabani, S. S., Faye, O., Pessôa, R., Patriota, J. V., Giorgi, R. R., et al. (2017). Rapid molecular detection of Zika virus in acute-phase urine samples using the recombinase polymerase amplification assay. PLoS Curr. 9:ecurrents.outbreaks.a7f1db2c7d66c3fc0ea0a774305d319e. doi: 10.1371/currents.outbreaks.a7f1db2c7d66c3fc0ea0a774305d319e

PubMed Abstract | CrossRef Full Text | Google Scholar

Acharya, D., Bastola, P., Le, L., Paul, A. M., Fernandez, E., Diamond, M. S., et al. (2016). An ultrasensitive electrogenerated chemiluminescence-based immunoassay for specific detection of Zika virus. Sci. Rep. 6:32227. doi: 10.1038/srep32227

PubMed Abstract | CrossRef Full Text | Google Scholar

Adegoke, O., Morita, M., Kato, T., Ito, M., Suzuki, T., and Park, E. Y. (2017). Localized surface plasmon resonance-mediated fluorescence signals in plasmonic nanoparticle-quantum dot hybrids for ultrasensitive Zika virus RNA detection via hairpin hybridization assays. Biosens. Bioelectron. 94, 513–522. doi: 10.1016/j.bios.2017.03.046

PubMed Abstract | CrossRef Full Text | Google Scholar

Al-Qahtani, A. A., Nazir, N., Al-Anazi, M. R., Rubino, S., and Al-Ahdal, M. N. (2016). Zika virus: a new pandemic threat. J. Infect. Dev. Ctries. 10, 201–207. doi: 10.3855/jidc.8350

PubMed Abstract | CrossRef Full Text | Google Scholar

Araujo Júnior, E., Carvalho, F. H., Tonni, G., and Werner, H. (2017). Prenatal imaging findings in fetal Zika virus infection. Curr. Opin. Obstet. Gynecol. 29, 95–105. doi: 10.1097/GCO.0000000000000345

PubMed Abstract | CrossRef Full Text | Google Scholar

Balm, M. N. D., Lee, C. K., Lee, H. K., Chiu, L., Koay, E. S. C., and Tang, J. W. (2012). A diagnostic polymerase chain reaction assay for Zika virus. J. Med. Virol. 84, 1501–1505. doi: 10.1002/jmv.23241

PubMed Abstract | CrossRef Full Text | Google Scholar

Balmaseda, A., Stettler, K., Medialdea-Carrera, R., Collado, D., Jin, X., Zambrana, J. V., et al. (2017). Antibody-based assay discriminates Zika virus infection from other flaviviruses. Proc. Natl. Acad. Sci U.S.A. 114, 8384–8389. doi: 10.1073/pnas.1704984114

PubMed Abstract | CrossRef Full Text | Google Scholar

Barzon, L., Pacenti, M., Berto, A., Sinigaglia, A., Franchin, E., Lavezzo, E., et al. (2016). Isolation of infectious Zika virus from saliva and prolonged viral RNA shedding in a traveller returning from the Dominican Republic to Italy, January 2016. Euro Surveill. 21, 1–5. doi: 10.2807/1560-7917.ES.2016.21.10.30159

PubMed Abstract | CrossRef Full Text | Google Scholar

Basile, A. J., Horiuchi, K., Panella, A. J., Laven, J., Kosoy, O., Lanciotti, R. S., et al. (2013). Multiplex microsphere immunoassays for the detection of IgM and IgG to arboviral diseases. PLoS ONE 8:e75670. doi: 10.1371/journal.pone.0075670

PubMed Abstract | CrossRef Full Text | Google Scholar

Bedin, F., Boulet, L., Voilin, E., Theillet, G., Rubens, A., and Rozand, C. (2017). Paper-based point-of-care testing for cost-effective diagnosis of acute flavivirus infections. J. Med. Virol. 89, 1520–1527. doi: 10.1002/jmv.24806

PubMed Abstract | CrossRef Full Text | Google Scholar

Bergquist, R. (2011). New tools for epidemiology: a space odyssey. Mem. Inst. Oswaldo Cruz. 106, 892–900. doi: 10.1590/S0074-02762011000700016

PubMed Abstract | CrossRef Full Text | Google Scholar

Bingham, A. M., Cone, M., Mock, V., Heberlein-Larson, L., Stanek, D., Blackmore, C., et al. (2016). Comparison of test results for Zika virus RNA in Urine, Serum, and Saliva specimens from persons with travel-associated Zika virus disease - Florida, 2016. MMWR Morb. Mortal. Wkly. Rep. 65, 475–478. doi: 10.15585/mmwr.mm6518e2

PubMed Abstract | CrossRef Full Text | Google Scholar

Boggild, A. K., Geduld, J., Libman, M., Yansouni, C. P., McCarthy, A. E., Hajek, J., et al. (2017). Surveillance report of Zika virus among Canadian travellers returning from the Americas. CMAJ. 189, E334–E340. doi: 10.1503/cmaj.161241

PubMed Abstract | CrossRef Full Text | Google Scholar

Bonaldo, M. C., Ribeiro, I. P., Lima, N. S., Dos Santos, A. A., Menezes, L. S., da Cruz, S. O., et al. (2016). Isolation of infective Zika virus from urine and Saliva of patients in Brazil. PLoS Negl. Trop. Dis. 10:e0004816. doi: 10.1371/journal.pntd.0004816

PubMed Abstract | CrossRef Full Text | Google Scholar

Bosch, I., de Puig, H., Hiley, M., Carré-Camps, M., Perdomo-Celis, F., Narváez, C. F., et al. (2017). Rapid antigen tests for dengue virus serotypes and Zika virus in patient serum. Sci. Transl. Med. 9:eaan1589. doi: 10.1126/scitranslmed.aan1589

PubMed Abstract | CrossRef Full Text | Google Scholar

Brault, A. C., and Bowen, R. A. (2016). The development of small animal models for Zika virus vaccine efficacy testing and pathological assessment. Am. J. Trop. Med. Hyg. 94, 1187–1188. doi: 10.4269/ajtmh.16-0277

PubMed Abstract | CrossRef Full Text | Google Scholar

Buechler, C. R., Bailey, A. L., Weiler, A. M., Barry, G. L., Breitbach, M. E., Stewart, L. M., et al. (2017). Seroprevalence of Zika virus in wild African green monkeys and baboons. mSphere. 2:e00392-16. doi: 10.1128/mSphere.00392-16

PubMed Abstract | CrossRef Full Text | Google Scholar

Calvet, G. A., Santos, F. B., and Sequeira, P. C. (2016). Zika virus infection: epidemiology, clinical manifestations and diagnosis. Curr. Opin. Infect. Dis. 29, 459–466. doi: 10.1097/QCO.0000000000000301

PubMed Abstract | CrossRef Full Text | Google Scholar

Cao-Lormeau, V. M., Blake, A., Mons, S., Lastère, S., Roche, C., Vanhomwegen, J., et al. (2016). Guillain-Barré Syndrome outbreak associated with Zika virus infection in French Polynesia: a case-control study. The Lancet. 387, 1531–1539. doi: 10.1016/S0140-6736(16)00562-6

PubMed Abstract | CrossRef Full Text | Google Scholar

Carod-Artal, F. J. (2016). Epidemiology and neurological complications of infection by the Zika virus: a new emerging neurotropic virus. Rev. Neurol. 62, 317–328.

PubMed Abstract | Google Scholar

Carteaux, G., Maquart, M., Bedet, A., Contou, D., Brugières, P., Fourati, S., et al. (2016). Zika virus associated with meningoencephalitis. N. Engl. J. Med. 374, 1595–1596. doi: 10.1056/NEJMc1602964

PubMed Abstract | CrossRef Full Text | Google Scholar

Chan, J. F., Yip, C. C., Tee, K. M., Zhu, Z., Tsang, J. O., Chik, K. K., et al. (2017). Improved detection of Zika virus RNA in human and animal specimens by a novel, highly sensitive and specific real-time RT-PCR assay targeting the 5'-untranslated region of Zika virus. Trop. Med. Int. Health. 22, 594–603. doi: 10.1111/tmi.12857

PubMed Abstract | CrossRef Full Text | Google Scholar

Chan, J. F.-W., Yip, C. C., Tsang, J. O., Tee, K. M., Cai, J. P., Chik, K. K., et al. (2016). Differential cell line susceptibility to the emerging Zika virus: implications for disease pathogenesis, non-vector-borne human transmission and animal reservoirs. Emerg. Microbes Infect. 5:e93. doi: 10.1038/emi.2016.99.

PubMed Abstract | CrossRef Full Text | Google Scholar

Chan, K., Weaver, S. C., Wong, P. Y., Lie, S., Wang, E., Guerbois, M., et al. (2016). Rapid, affordable and portable medium-throughput molecular device for Zika virus. Sci. Rep. 6:38223. doi: 10.1038/srep38223

PubMed Abstract | CrossRef Full Text | Google Scholar

Chotiwan, N., Brewster, C. D., Magalhaes, T., Weger-Lucarelli, J., Duggal, N. K., Rückert, C., et al. (2017). Rapid and specific detection of Asian- and African-lineage Zika viruses. Sci. Transl. Med. 9:eaag0538. doi: 10.1126/scitranslmed.aag0538

PubMed Abstract | CrossRef Full Text | Google Scholar

Coelho, S. V. A., Neris, R. L. S., Papa, M. P., Schnellrath, L. C., Meuren, L. M., Tschoeke, D. A., et al. (2017). Development of standard methods for Zika virus propagation, titration, and purification. J. Virol. Methods. 246, 65–74. doi: 10.1016/j.jviromet.2017.04.011

PubMed Abstract | CrossRef Full Text | Google Scholar

Cordeiro, M. T. (2017). “Zika virus: laboratory diagnosis,” in Zika in Focus, ed M. Vasco Aragão (Cham: Springer), 9–62.

Google Scholar

Corman, V. M., Rasche, A., Baronti, C., Aldabbagh, S., Cadar, D., Reusken, C. B. E. M., et al. (2016). Assay optimization for molecular detection of Zika virus. Bull. World Health Organ. 94, 880–892. doi: 10.2471/BLT.16.175950

PubMed Abstract | CrossRef Full Text | Google Scholar

Cortese, M., Goellner, S., Acosta, E. G., Neufeldt, C. J., Oleksiuk, O., Lampe, M., et al. (2017). Ultrastructural characterization of Zika virus replication factories. Cell Rep. 18, 2113–2123. doi: 10.1016/j.celrep.2017.02.014

PubMed Abstract | CrossRef Full Text | Google Scholar

de Fatima Vasco Aragao, M., van der Linden, V., Brainer-Lima, A. M., Coeli, R. R., Rocha, M. A., Sobral da Silva, P., et al. (2016). Clinical features and neuroimaging (CT and MRI) findings in presumed Zika virus related congenital infection and microcephaly: retrospective case series study. BMJ. 353, i1901. doi: 10.1136/bmj.i3182

PubMed Abstract | CrossRef Full Text

de M. Campos, R., Cirne-Santos, C., Meira, G. L. S., Santos, L. L. R., de Meneses, M. D., Friedrich, J., et al. (2016). Prolonged detection of Zika virus RNA in urine samples during the ongoing Zika virus epidemic in Brazil. J. Clin. Virol. 77, 69–70. doi: 10.1016/j.jcv.2016.02.009

CrossRef Full Text | Google Scholar

de Oliveira-Szejnfeld, P. S., Levine, D., Melo, A. S. D. O., Amorim, M. M. R., Batista, A. G. M., Chimelli, L., et al. (2017). Congenital brain abnormalities and Zika virus: what the radiologist can expect to see prenatally and postnatally. Radiology 281, 203–218. doi: 10.1148/radiol.2016161584

CrossRef Full Text | Google Scholar

Dhama, K., Karthik, K., Chakraborty, S., Tiwari, R., Kapoor, S., Kumar, A., et al. (2014). Loop-mediated isothermal amplification of DNA (LAMP) – a new diagnostic tool lights the world of diagnosis of animal and human pathogens: a review. Pak. J. Biol. Sci. 17, 151–166. doi: 10.3923/pjbs.2014.151.166

PubMed Abstract | CrossRef Full Text | Google Scholar

Dhama, K., Verma, A. K., Tiwari, R., Chakraborty, S., Vora, K., Kapoor, S., et al. (2013). A perspective on applications of geographical information system (GIS); an advanced tracking tool for disease surveillance and monitoring in veterinary epidemiology. Adv. Anim. Vet. Sci. 1, 14–24.

Eboigbodin, K. E., Brummer, M., Ojalehto, T., and Hoser, M. (2016). Rapid molecular diagnostic test for Zika virus with low demands on sample preparation and instrumentation. Diag. Microb. Infect. Dis. 86, 369–371. doi: 10.1016/j.diagmicrobio.2016.08.027

PubMed Abstract | CrossRef Full Text | Google Scholar

Eppes, C., Rac, M., Dunn, J., Versalovic, J., Murray, K. O., Suter, M. A., et al. (2017). Testing for Zika virus infection in pregnancy: key concepts to deal with an emerging epidemic. Am. J. Obstet. Gynecol. 216, 209–225. doi: 10.1016/j.ajog.2017.01.020

PubMed Abstract | CrossRef Full Text | Google Scholar

Fagbami, A. H. (1979). Zika virus infections in Nigeria: virological and seroepidemiological investigations in Oyo State. J. Hyg. 83, 213–219. doi: 10.1017/S0022172400025997

PubMed Abstract | CrossRef Full Text | Google Scholar

Fajardo, A., Cristina, J., and Moreno, P. (2016). Emergence and spreading potential of Zika virus. Front. Microbiol. 7:1667. doi: 10.3389/fmicb.2016.01667

PubMed Abstract | CrossRef Full Text | Google Scholar

Faye, O., Faye, O., Diallo, D., Diallo, M., Weidmann, M., and Sall, A. A. (2013). Quantitative real-time PCR detection of Zika virus and evaluation with field-caught mosquitoes. Virol. J. 10, 311–318. doi: 10.1186/1743-422X-10-311

PubMed Abstract | CrossRef Full Text | Google Scholar

Faye, O., Faye, O., Dupressoir, A., Weidmann, M., Ndiaye, M., and Alpha Sall, A. (2008). One-step RT-PCR for detection of Zika virus. J. Clin. Virol. 43, 96–101. doi: 10.1016/j.jcv.2008.05.005

PubMed Abstract | CrossRef Full Text | Google Scholar

FDA (2016). Zika Virus Emergency Use Authorization. Silver Spring, MD: US Department of Health and Human Services, Food and Drug Administration. Available online at: http://www.fda.gov/MedicalDevices/Safety/EmergencySituations/ucm161496.htm

Fonseca, K., Meatherall, B., Zarra, D., Drebot, M., MacDonald, J., Pabbaraju, K., et al. (2014). First case of Zika virus infection in a returning Canadian traveler. Am. J. Trop. Med. Hyg. 91, 1035–1038. doi: 10.4269/ajtmh.14-0151

PubMed Abstract | CrossRef Full Text | Google Scholar

Frankel, M. B., Pandya, K., Gersch, J., Siddiqui, S., and Schneider, G. J. (2017). Development of the Abbott RealTime ZIKA assay for the qualitative detection of Zika virus RNA from serum, plasma, urine, and whole blood specimens using the m2000 system. J. Virol. Methods 246, 117–124. doi: 10.1016/j.jviromet.2017.05.002

PubMed Abstract | CrossRef Full Text | Google Scholar

Fung, I. C., Blankenship, E. B., Goff, M. E., Mullican, L. A., Chan, K. C., Saroha, N., et al. (2017). Zika-virus-related photo sharing on Pinterest and Instagram. Disaster Med. Public Health Prep. 23, 1–4. doi: 10.1017/dmp.2017.23

CrossRef Full Text | Google Scholar

Gourinat, A. C., O'Connor, O., Calvez, E., Goarant, C., and Dupont-Rouzeyrol, M. (2015). Detection of Zika virus in urine. Emerging Infect. Dis. 21, 84–86. doi: 10.3201/eid2101.140894

PubMed Abstract | CrossRef Full Text | Google Scholar

Granger, D., Hilgart, H., Misner, L., Christensen, J., Bistodeau, S., Palm, J., et al. (2017). Serologic testing for zika virus: comparison of three Zika virus igm elisas and initial laboratory experiences. J. Clin. Microbiol. 55, 2127–2136. doi: 10.1128/JCM.00580-17

CrossRef Full Text | Google Scholar

Grard, G., Caron, M., Mombo, I. M., Nkoghe, D., Mboui Ondo, S., Jiolle, D., et al. (2014). Zika virus in Gabon (Central Africa) 2007: a new threat from Aedes albopictus? PLoS Negl. Trop. Dis. 8:e2681. doi: 10.1371/journal.pntd.0002681

PubMed Abstract | CrossRef Full Text | Google Scholar

Gu, S. H., Song, D. H., Lee, D., Jang, J., Kim, M. Y., Jung, J., et al. (2017). Whole-genome sequence analysis of Zika virus, amplified from urine of traveler from the Philippines. Virus Genes 53, 918–921. doi: 10.1007/s11262-017-1500-9

PubMed Abstract | CrossRef Full Text | Google Scholar

Haddow, A. J., Williams, M. C., Woodall, J. P., Simpson, D. I., and Goma, L. K. (1964). Twelve isolations of Zika virus from Aedes (Stegomyia) africanus (theobald) taken in and above a Uganda forest. Bull. World Health Organ. 31, 57–69.

PubMed Abstract | Google Scholar

Hamel, R., Dejarnac, O., Wichit, S., Ekchariyawat, P., Neyret, A., Luplertlop, N., et al. (2015). Biology of Zika virus infection in human skin cell. J. Virol. 89, 8880–8896. doi: 10.1128/JVI.00354-15

CrossRef Full Text | Google Scholar

Hamel, R., Liégeois, F., Wichit, S., Pompon, J., Diop, F., Talignani, L., et al. (2016). Zika virus: epidemiology, clinical features and host-virus interactions. Microbes Infect. 18, 441–449. doi: 10.1016/j.micinf.2016.03.009

PubMed Abstract | CrossRef Full Text | Google Scholar

Hancock, W. T., Soeters, H. M., Hills, S. L., Link-Gelles, R., Evans, M. E., Daley, W. R., et al. (2017). Establishing a timeline to discontinue routine testing of asymptomatic pregnant women for Zika virus infection - American samoa, 2016-2017. MMWR Morb. Mortal. Wkly. Rep. 66, 299–301. doi: 10.15585/mmwr.mm6611a5

PubMed Abstract | CrossRef Full Text | Google Scholar

Haug, C. J., Kieny, M. P., and Murgue, B. (2016). The Zika challenge. N. Engl. J. Med. 374, 1801–1803. doi: 10.1056/NEJMp1603734

PubMed Abstract | CrossRef Full Text | Google Scholar

Hayes, E. B. (2009). Zika virus outside Africa. Emerg. Infect. Dis. 15, 1347–1350. doi: 10.3201/eid1509.090442

PubMed Abstract | CrossRef Full Text | Google Scholar

Hills, S. L., Russell, K., Hennessey, M., Williams, C., Oster, A. M., Fischer, M., et al. (2016). Transmission of Zika virus through sexual contact with travelers to areas of ongoing transmission–continental United States, 2016. MMWR Morb. Mortal. Wkly. Rep. 65, 215–216. doi: 10.15585/mmwr.mm6508e2

PubMed Abstract | CrossRef Full Text | Google Scholar

Ioos, S., Mallet, H. P., Leparc Goffart, I., Gauthier, V., Cardoso, T., and Herida, M. (2014). Current Zika virus epidemiology and recent epidemics. Med. Mal. Infect. 44, 302–307. doi: 10.1016/j.medmal.2014.04.008

PubMed Abstract | CrossRef Full Text | Google Scholar

Jacob, J. A. (2016). Researchers focus on solving the Zika riddles. JAMA. 315, 1097–1099. doi: 10.1001/jama.2016.1219

PubMed Abstract | CrossRef Full Text | Google Scholar

Jiang, Q., Chandar, Y. J., Cao, S., Kharasch, E. D., Singamaneni, S., and Morrissey, J. J. (2017). Rapid, point-of-care, paper-based plasmonic biosensor for Zika virus diagnosis. Adv. Biosys. 1:1700096. doi: 10.1002/adbi.201700096

CrossRef Full Text | Google Scholar

Kadkhoda, K., Gretchen, A., and Racano, A. (2017). Evaluation of a commercially available Zika virus IgM ELISA: specificity in focus. Diagn. Microbiol. Infect. Dis. 88, 233–235. doi: 10.1016/j.diagmicrobio.2017.04

PubMed Abstract | CrossRef Full Text | Google Scholar

Karthik, K., Rathore, R., Thomas, P., Arun, T. R., Viswas, K. N., Dhama, K., et al. (2014). New closed tube loop mediated isothermal amplification assay for prevention of product cross contamination. MethodsX 1, e137–e143. doi: 10.1016/j.mex.2014.08.009

PubMed Abstract | CrossRef Full Text | Google Scholar

Kawadkar, J., Chauhan, M. K., and Maharana, M. (2011). Nanobiotechnology: application of nanotechnology in diagnosis, drug discovery and drug development. Asian Pharmaceut Clin Res. 4, 23–28.

Google Scholar

Kelvin, A. A., Banner, D,., Pamplona, L., Alencar, C., Rubino, S., and Heukelbach, J. (2016). ZIKATracker: a mobile App for reporting cases of ZIKV worldwide. J. Infect. Dev. Ctries. 10, 113–115. doi: 10.3855/jidc.8248

PubMed Abstract | CrossRef Full Text | Google Scholar

Khandia, R., Munjal, A., and Dhama, K. (2017). Consequences of Zika virus infection during fetal stage and pregnancy safe drugs: an update. Int. J. Pharmacol. 13, 370–377. doi: 10.3923/ijp.2017.370.377

CrossRef Full Text

Kirya, B. G., and Okia, N. O. (1977). A yellow fever epizootic in Zika Forest, Uganda, during 1972: Part 2: monkey serology. Trans. R. Soc. Trop. Med. Hyg. 71, 300–303. doi: 10.1016/0035-9203(77)90104-3

PubMed Abstract | CrossRef Full Text | Google Scholar

Koff, W. C., Burton, D. R., Johnson, P. R., Walker, B. D., King, C. R., Nabel, G. J., et al. (2013). Accelerating next generation vaccine development for global disease prevention. Science 340:1232910. doi: 10.1126/science.1232910

PubMed Abstract | CrossRef Full Text | Google Scholar

Koide, F., Goebel, S., Snyder, B., Walters, K. B., Gast, A., Hagelin, K., et al. (2016). Development of a Zika virus infection model in cynomolgus macaques. Front. Microbiol. 7:2028. doi: 10.3389/fmicb.2016.02028

PubMed Abstract | CrossRef Full Text | Google Scholar

Kruger, R. P. (2016). Zika virus on the move. Cell 164, 585–587. doi: 10.1016/j.cell.2016.01.040

CrossRef Full Text

Lamb, L. E., Bartolone, S. N., Kutluay, S. B., Robledo, D., Porras, A., Plata, M., et al. (2016). Advantage of urine based molecular diagnosis of Zika virus. Int. Urol. Nephrol. 48, 1961–1966. doi: 10.1007/s11255-016-1406-9

PubMed Abstract | CrossRef Full Text | Google Scholar

Lambe, U., Prasad, M., Brar, B., Guray, M., Ikbal Ranjan, K., et al. (2016). Nanodiagnostics: a new frontier for veterinary and medical sciences. J. Exp. Biol. Agric. Sci. 4.307–320. doi: 10.18006/2016.4(3S).307.320

CrossRef Full Text | Google Scholar

Lanciotti, R. S., Kosoy, O. L., Laven, J. J., Velez, J. O., Lambert, A. J., Johnson, A. J., et al. (2008). Genetic and serologic properties of Zika virus associated with an epidemic, Yap state, Micronesia, 2007. Emerg. Infect. Dis. 14, 1232–1239. doi: 10.3201/eid1408.080287

PubMed Abstract | CrossRef Full Text | Google Scholar

Landry, M. L., and St. George, K. (2017). Laboratory diagnosis of Zika virus infection. Arch. Pathol. Lab. Med. 141, 60–67. doi: 10.5858/arpa.2016-0406-SA

PubMed Abstract | CrossRef Full Text | Google Scholar

Lazear, H. M., and Diamond, M. S. (2016). Zika virus: new clinical syndromes and its emergence in the Western Hemisphere. J. Virol. 90, 4864–4875 doi: 10.1128/JVI.00252-16

PubMed Abstract | CrossRef Full Text | Google Scholar

Lee, D., Shin, Y., Chung, S., Hwang, K. S., Yoon, D. S., and Lee, J. H. (2016). Simple and highly sensitive molecular diagnosis of Zika virus by lateral flow assays. Anal. Chem. 88, 12272–12278. doi: 10.1021/acs.analchem.6b03460

PubMed Abstract | CrossRef Full Text | Google Scholar

Leung, G. H., Baird, R. W., Druce, J., and Anstey, N. M. (2015). Zika virus infection in Australia following a monkey bite in Indonesia. Southeast Asian J. Trop. Med. Public Health. 46, 460–464.

PubMed Abstract | Google Scholar

L'Huillier, A. G., Hamid-Allie, A., Kristjanson, E., Papageorgiou, L., Hung, S., Wong, C. F., et al. (2017a). Evaluation of Euroimmun anti-Zika virus IgM and IgG enzyme-linked immunosorbent assays for Zika virus serologic testing. J. Clin. Microbiol. 55, 2462–2471. doi: 10.1128/JCM.00442-17

PubMed Abstract | CrossRef Full Text

L'Huillier, A. G., Lombos, E., Tang, E., Perusini, S., Eshaghi, A., Nagra, S., et al. (2017b). Evaluation of altona diagnostics realstar Zika virus RT-PCR test kit for Zika virus PCR testing. J. Clin. Microbiol. 55, 1576–1584. doi: 10.1128/JCM.02153-16

CrossRef Full Text | Google Scholar

Lum, F. M., Lin, C., Susova, O. Y., Teo, T. H., Fong, S. W., Mak, T. M., et al. (2017). Sensitive detection of Zika virus antigen in patients' whole blood as an alternative diagnostic approach. J. Infect. Dis. 216, 182–190. doi: 10.1093/infdis/jix276

CrossRef Full Text | Google Scholar

Lustig, Y., Zelena, H., Venturi, G., Van Esbroeck, M., Rothe, C., Perret, C., et al. (2017). Sensitivity and kinetics of a NS1-based Zika virus ELISA in Zika infected travelers from Israel, Czech Republic, Italy, Belgium, Germany and Chile. J. Clin. Microbiol. 55, 1894–1901. doi: 10.1128/JCM.00346-17

CrossRef Full Text | Google Scholar

Marchette, N. J., Garcia, R., and Rudnick, A. (1969). Isolation of Zika virus from Aedes aegypti mosquitoes in Malaysia. Am. J. Trop. Med. Hyg. 18, 411–415. doi: 10.4269/ajtmh.1969.18.411

PubMed Abstract | CrossRef Full Text | Google Scholar

Martines, R. B., Bhatnagar, J., Keating, M. K., Silva-Flannery, L., Muehlenbachs, A., Gary, J., et al. (2016). Notes from the Field: evidence of Zika virus infection in brain and placental tissues from two congenitally infected newborns and two fetal losses - Brazil, 2015. MMWR Morb. Mortal. Wkly. Rep. 65, 159–160. doi: 10.15585/mmwr.mm6506e1

PubMed Abstract | CrossRef Full Text | Google Scholar

Metsky, H. C., Matranga, C. B., Wohl, S., Schaffner, S. F., Freije, C. A., Winnicki, S. M., et al. (2017). Zika virus evolution and spread in the Americas. Nature 546, 411–415. doi: 10.1038/nature22402

PubMed Abstract | CrossRef Full Text | Google Scholar

Miller, E., Becker, Z., Shalev, D., Lee, C. T., Cioroiu, C., and Thakur, K. (2017). Probable Zika virus-associated Guillain-Barré syndrome: challenges with clinico-laboratory diagnosis. J. Neurol. Sci. 375, 367–370. doi: 10.1016/j.jns.2017.02.029

PubMed Abstract | CrossRef Full Text | Google Scholar

Mlakar, J., Korva, M., Tul, N., Popović, M., Poljšak-Prijatelj, M., Mraz, J., et al. (2016). Zika virus associated with microcephaly. N. Engl. J. Med. 374, 951–958. doi: 10.1056/NEJMoa1600651

PubMed Abstract | CrossRef Full Text

Monath, T. P., Craven, R. B., Muth, D. J., Trautt, C. J., Calisher, C. H., and Fitzgerald, S. A. (1980). Limitations of the complement-fixation test for distinguishing naturally acquired from vaccine induced yellow fever infection in flavivirus-hyperendemic areas. Am. J. Trop. Med. Hyg. 29, 624–634. doi: 10.4269/ajtmh.1980.29.624

PubMed Abstract | CrossRef Full Text | Google Scholar

Morrissey, J., Jiang, Q., Chandar, Y., Cao, S., Kharasch, E., and Singamaneni, S. (2017). “Rapid, point-of-care, thermally stable paper-based plasmonic assay for Zika virus diagnosis,” in Advanced Photonics 2017 (IPR, NOMA, Sensors, Networks, SPPCom, PS), OSA Technical Digest (online) (Optical Society of America, 2017), paper SeTu1E.2.

Google Scholar

Moulin, E., Selby, K., Cherpillod, P., Kaiser, L., and Boillat-Blanco, N. (2016). Simultaneous outbreaks of dengue, chikungunya and Zika virus infections: diagnosis challenge in a returning traveller with nonspecific febrile illness. New Microbes New Infect. 11, 6–7. doi: 10.1016/j.nmni.2016.02.003

PubMed Abstract | CrossRef Full Text | Google Scholar

Müller, J. A., Harms, M., Schubert, A., Mayer, B., Jansen, S., Herbeuval, J. P., et al. (2017). Development of a high-throughput colorimetric Zika virus infection assay. Med. Microbiol. Immunol. 206, 175–185. doi: 10.1007/s00430-017-0493-2

PubMed Abstract | CrossRef Full Text | Google Scholar

Munjal, A., Khandia, R., Dhama, K., Sachan, S., Karthik, K., Tiwari, R., et al. (2017a). Advances in developing therapies to combat Zika virus: current knowledge and future perspectives. Front. Microbiol. 8:1469. doi: 10.3389/fmicb.2017.01469

PubMed Abstract | CrossRef Full Text | Google Scholar

Munjal, A., Khandia, R., Tiwari, R., Chakraborty, S., Karthik, K., and Dhama, K. (2017b). Advances in designing and developing vaccines against Zika virus. Int. J. Pharmacol. 13, 667–676. doi: 10.3923/ijp.2017.667.676

CrossRef Full Text

Murray, K. O., Gorchakov, R., Carlson, A. R., Berry, R., Lai, L., Natrajan, M., et al. (2017). Prolonged detection of zika virus in vaginal secretions and whole blood. Emerg. Infect. Dis. 23, 99–101. doi: 10.3201/eid2301.161394

PubMed Abstract | CrossRef Full Text | Google Scholar

Musso, D., and Gubler, D. J. (2016). Zika virus. Clin. Microbiol. Rev. 29, 487–524. doi: 10.1128/CMR.00072-15

PubMed Abstract | CrossRef Full Text | Google Scholar

Musso, D., Roche, C., Nhan, T. X., Robin, E., Teissier, A., and Cao-Lormeau, VM. (2015). Detection of Zika virus in saliva. J. Clin. Virol. 68, 53–55. doi: 10.1016/j.jcv.2015.04.021

PubMed Abstract | CrossRef Full Text | Google Scholar

Nicolini, A. M., McCracken, K. E., and Yoon, J. Y. (2017). Future developments in biosensors for field-ready Zika virus diagnostics. J. Biol. Eng. 11:7. doi: 10.1186/s13036-016-0046-z

PubMed Abstract | CrossRef Full Text | Google Scholar

Nikolay, A., Castilho, L. R., Reichl, U., and Genzel, Y. (in press). Propagation of Brazilian Zika virus strains in static suspension cultures using Vero BHK cells. Vaccine. doi: 10.1016/j.vaccine.2017.03.018

PubMed Abstract | CrossRef Full Text | Google Scholar

Nishiura, H., Kinoshita, R., Mizumoto, K., Yasuda, Y., and Nah, K. (2016). Equation Title: transmission potential of Zika virus infection in the South Pacific. Int. J. Infect. Dis. 45, 95–97. doi: 10.1016/j.ijid.2016.02.017

CrossRef Full Text | Google Scholar

Novotny, J., Svobodova, J., Ransnas, L. A., and Kubistova, K. (1992). A method for the preparation of purified antigens of coxsackie virus B3 from a large volume of cell culture supernatant. Acta Virol. 36, 483–487.

Oliveira Melo, A. S., Malinger, G., Ximenes, R., Szejnfeld, P. O., Alves Sampaio, S., and Bispo de Filippis, A. M. (2016). Zika virus intrauterine infection causes fetal brain abnormality and microcephaly: tip of the iceberg? Ultrasound Obstet. Gynecol. 47, 6–7. doi: 10.1002/uog.15831

PubMed Abstract | CrossRef Full Text | Google Scholar

Pardee, K., Green, A. A., Takahashi, M. K., Braff, D., Lambert, G., Lee, J. W., et al. (2016). Rapid, low cost detection of Zika virus using programmable biomolecular components. Cell 165, 1255–1266. doi: 10.1016/j.cell.2016.04.059

PubMed Abstract | CrossRef Full Text | Google Scholar

Pardy, R. D., Rajah, M. M., Condotta, S. A., Taylor, N. G., Sagan, S. M., and Richer, M. J. (2017). Analysis of the T Cell response to Zika virus and identification of a novel CD8+ T cell epitope in immunocompetent mice. PLoS Pathog. 13:e1006184. doi: 10.1371/journal.ppat.1006184

PubMed Abstract | CrossRef Full Text | Google Scholar

Parra-Saavedra, M., Reefhuis, J., Piraquive, J. P., Gilboa, S. M., Badell, M. L., et al. (2017). Serial head and brain imaging of 17 fetuses with confirmed Zika virus infection in Colombia, South America. Obstet. Gynecol. 130, 207–212. doi: 10.1097/AOG.0000000000002105

PubMed Abstract | CrossRef Full Text | Google Scholar

Pasquier, C., Joguet, G., Mengelle, C., Chapuy-Regaud, S., Pavili, L., Prisant, N., et al. (2018). Kinetics of anti-ZIKV antibodies after Zika infection using two commercial enzyme-linked immunoassays. Diagn. Microbiol. Infect. Dis. 90, 26–30. doi: 10.1016/j.diagmicrobio.2017.09.001

PubMed Abstract | CrossRef Full Text | Google Scholar

Pessôa, R., Patriota, J. V., de Souza, Mde L., Abd El Wahed, A., and Sanabani, S. S. (2016). Detection of Zika virus in Brazilian patients during the first five days of infection - urine versus plasma. Euro Surveill. 21:30302. doi: 10.2807/1560-7917.ES.2016.21.30.30302

PubMed Abstract | CrossRef Full Text | Google Scholar

Petersen, L. R., Jamieson, D. J., Powers, A. M., and Honein, M. A. (2016). Zika virus. N. Engl. J. Med. 374, 1552–1563. doi: 10.1056/NEJMra1602113

CrossRef Full Text | Google Scholar

Pinto, V. L. Jr., Luz, K., Parreira, R., and Ferrinho, P. (2015). Zika virus: a review to clinicians. Acta Med. Port. 28, 760–765.

Google Scholar

Priye, A., Bird, S. W., Light, Y. K., Ball, C. S., Negrete, O. A., and Meagher, R. J. (2017). A smart phone-based diagnostic platform for rapid detection of Zika, chikungunya, and dengue viruses. Sci. Rep. 7:44778. doi: 10.1038/srep44778

CrossRef Full Text | Google Scholar

Pyzocha, N. J., Chinchen, S. E., and Maurer, D. M. (2017). Zany over Zika virus: an overview of diagnosis and treatment modalities. Curr. Sports Med. Rep. 16, 109–113. doi: 10.1249/JSR.0000000000000346

PubMed Abstract | CrossRef Full Text | Google Scholar

Quick, J., Grubaugh, N. D., Pullan, S. T., Claro, I. M., Smith, A. D., Gangavarapu, K., et al. (2017). Multiplex PCR method for MinION and Illumina sequencing of Zika and other virus genomes directly from clinical samples. Nat. Protoc. 12, 1261–1276. doi: 10.1038/nprot.2017.066

PubMed Abstract | CrossRef Full Text | Google Scholar

Rafiei, N., Hajkowicz, K., Redmond, A., and Taylor, C. (2016). First report of Zika virus infection in a returned traveller from the Solomon Islands. Med. J. Aust. 204, 186. doi: 10.5694/mja15.01275

PubMed Abstract | CrossRef Full Text | Google Scholar

Ratcliff, R. M., Chang, G., Kok, T., and Sloots, T. P. (2007). Molecular diagnosis of medical viruses. Curr Iss Molr Biol. 9, 87–102.

PubMed Abstract | Google Scholar

Rather, I. A., Kumar, S., Bajpai, V. K., Lim, J., and Park, YH. (2017). Prevention and control strategies to counter Zika epidemic. Front. Microbiol. 8:305. doi: 10.3389/fmicb.2017.00305

PubMed Abstract | CrossRef Full Text | Google Scholar

Ren, P., Ortiz, D. A., Terzian, A. C. B., Colombo, T. E., Nogueira, M. L., Vasilakis, N., et al. (2017). Evaluation of Aptima Zika virus assay. J. Clin. Microbiol. 55, 2198–2203. doi: 10.1128/JCM.00603-17

PubMed Abstract | CrossRef Full Text | Google Scholar

Rodriguez-Morales, A. J., Galindo-Marquez, M. L., García-Loaiza, C. J., Sabogal-Roman, J. A., Marin-Loaiza, S., Ayala, A. F., et al. (2016). Mapping Zika virus infection using geographical information systems in Tolima, Colombia, 2015-2016. F1000Res. 5:568. doi: 10.12688/f1000research.8436.1

PubMed Abstract | CrossRef Full Text | Google Scholar

Rodriguez-Morales, A. J., Galindo-Marquez, M. L., García-Loaiza, C. J., Sabogal-Roman, J. A., Marin-Loaiza, S., Ayala, A. F., Lagos-Grisales, G. J., et al. (2017a). Mapping Zika virus disease incidence in Valle del Cauca. Infection 45, 93–102. doi: 10.1007/s15010-016-0948-1

PubMed Abstract | CrossRef Full Text | Google Scholar

Rodriguez-Morales, A. J., Ruiz, P., Tabares, J., Ossa, C. A., Yepes-Echeverry, M. C., Ramirez-Jaramillo, V., et al. (2017b). Mapping the ecoepidemiology of Zika virus infection in urban and rural areas of Pereira, Risaralda, Colombia, 2015-2016: implications for public health and travel medicine. Travel Med. Infect. Dis. 18, 57–66. doi: 10.1016/j.tmaid.2017.05.004

PubMed Abstract | CrossRef Full Text | Google Scholar

Rossi, S. L., Tesh, R. B., Azar, S. R., Muruato, A. E., Hanley, K. A., Auguste, A. J., et al. (2016). Characterization of a novel murine model to study Zika virus. Am. J. Trop. Med. Hyg. 94, 1362–1369. doi: 10.4269/ajtmh.16-0111

PubMed Abstract | CrossRef Full Text | Google Scholar

Rossini, G., Gaibani, P., Vocale, C., Cagarelli, R., and Landini, M. P. (2017). Comparison of Zika virus (ZIKV) RNA detection in plasma, whole blood and urine - case series of travel-associated ZIKV infection imported to Italy, 2016. J. Infect. 75, 242–245. doi: 10.1016/j.jinf.2017.05.021

PubMed Abstract | CrossRef Full Text | Google Scholar

Rozé, B., Najioullah, F., Fergé, J. L., Apetse, K., Brouste, Y., Cesaire, R., et al. (2016). Zika virus detection in urine from patients with Guillain-Barré syndrome on Martinique. Euro Surveill. 21:30154. doi: 10.2807/1560-7917.ES.2016.21.9.30154

PubMed Abstract | CrossRef Full Text | Google Scholar

Sakudo, A., Viswan, A., Chou, H., Sasaki, T., Ikuta, K., and Nagatsu, M. (2016). Capture of Dengue viruses using antibody-integrated graphite-encapsulated magnetic beads produced using gas plasma technology. Mol. Med. Rep. 14, 697–704. doi: 10.3892/mmr.2016.5330

PubMed Abstract | CrossRef Full Text | Google Scholar

Sánchez-Purrà, M., Carré-Camps, M., de Puig, H., Bosch, I., Gehrke, L., and Hamad-Schifferli, K. (2017). Surface-enhanced Raman spectroscopy based sandwich immunoassays for multiplexed detection of zika and dengue viral biomarkers. ACS Infect Dis. 3, 767–776. doi: 10.1021/acsinfecdis.7b00110.

PubMed Abstract | CrossRef Full Text | Google Scholar

Saraswati, T. E., Ogino, A., and Nagatsu, M. (2012). Plasma-activated immobilization of biomolecules on to graphite-encapsulated magnetic nanoparticles. Carbon N.Y. 50, 1253–1261. doi: 10.1016/j.carbon.2011.10.044

CrossRef Full Text | Google Scholar

Shadgan, B., Pakravan, A., Zaeimkohan, H., Shahpar, F. M., and Khodaee, M. (2016). Zika and Rio Olympic games. Curr. Sports Med. Rep. 15, 298–300. doi: 10.1249/JSR.0000000000000278

PubMed Abstract | CrossRef Full Text | Google Scholar

Shan, C., Ortiz, D. A., Yang, Y., Wong, S. J., Kramer, L. D., Shi, P.-Y., et al. (2017b). Evaluation of a novel reporter virus neutralization test for the serological diagnosis of Zika and dengue virus infection. J. Clin. Microbiol. doi: 10.1128/JCM.00975-17

PubMed Abstract | CrossRef Full Text | Google Scholar

Shan, C., Xie, X., Ren, P., Loeffelholz, M. J., Yang, Y., Furuya, A., et al. (2017a). A rapid Zika diagnostic assay to measure neutralizing antibodies in patients. EBioMedicine 17, 157–162. doi: 10.1016/j.ebiom.2017.03.006

PubMed Abstract | CrossRef Full Text | Google Scholar

Shankar, A., Patil, A. A., and Skariyachan, S. (2017). Recent perspectives on genome, transmission, clinical manifestation, diagnosis, therapeutic strategies, vaccine developments, and challenges of Zika virus research. Front. Microbiol. 8:1761. doi: 10.3389/fmicb.2017.01761

PubMed Abstract | CrossRef Full Text | Google Scholar

Sharma, A., and Lal, S. K. (2017). Zika virus: transmission, detection, control, and prevention. Front. Microbiol. 8:110. doi: 10.3389/fmicb.2017.00110

PubMed Abstract | CrossRef Full Text | Google Scholar

Shinohara, K., Kutsuna, S., Takasaki, T., Moi, M. L., Ikeda, M., Kotaki, A., et al. (2016). Zika fever imported from Thailand to Japan, and diagnosed by PCR in the urines. J. Travel. Med. 23:tav011. doi: 10.1093/jtm/tav011

PubMed Abstract | CrossRef Full Text | Google Scholar

Shukla, S., Hong, S. Y., Chung, S. H., and Kim, M. (2016). Rapid detection strategies for the global threat of Zika virus: current state, new hypotheses, and limitations. Front. Microbiol. 7:1685. doi: 10.3389/fmicb.2016.01685

PubMed Abstract | CrossRef Full Text | Google Scholar

Singh, R. K., Badasara, S. K., Dhama, K., and Malik, Y. P. S. (2015). “Progress and prospects in vaccine research,” in National Workshop on “Current Trends and Future Research Challenges in Vaccines and Adjuvants” (Bareilly: Organized at ICAR Indian Veterinary Research Institute), 1–19.

Singh, R. K., Dhama, K., Malik, Y. S., Ramakrishnan, M. A., Karthik, K., Khandia, R., et al. (2017). Ebola virus – epidemiology, diagnosis and control: threat to humans, lessons learnt, and preparedness plans- an update on its 40 year's journey. Vet. Quart 37, 98–135. doi: 10.1080/01652176.2017.1309474

PubMed Abstract | CrossRef Full Text | Google Scholar

Singh, R. K., Dhama, K., Malik, Y. S., Ramakrishnan, M. A., Karthik, K., Tiwari, R., et al. (2016). Zika virus – emergence, evolution, pathology, diagnosis, and control: current global scenario and future perspectives – a comprehensive review. Vet. Quart. 36, 150–175. doi: 10.1080/01652176.2016.1188333

PubMed Abstract | CrossRef Full Text | Google Scholar

Society for Maternal-Fetal Medicine (SMFM) Publications Committee (2016). Ultrasound screening for fetal microcephaly following Zika virus exposure. Am. J. Obstet. Gynecol. 214, B2–4. doi: 10.1016/j.ajog.2016.02.043

CrossRef Full Text

Song, J., Mauk, M. G., Hackett, B. A., Cherry, S., Bau, H. H., and Liu, C. (2016). Instrument-free point-of-care molecular detection of Zika virus. Anal. Chem. 88, 7289–7294. doi: 10.1021/acs.analchem.6b01632

PubMed Abstract | CrossRef Full Text | Google Scholar

Staples, J. E., Dziuban, E. J., Fischer, M., Cragan, J. D., Rasmussen, S. A., Cannon, M. J., et al. (2016). Interim guidelines for the evaluation and testing of infants with possible congenital Zika virus infection United States. MMWR Morb. Mortal. Wkly. Rep. 65, 63–67. doi: 10.15585/mmwr.mm6503e3

PubMed Abstract | CrossRef Full Text | Google Scholar

Steinhagen, K., Probst, C., Radzimski, C., Schmidt-Chanasit, J., Emmerich, P., van Esbroeck, M., et al. (2016). Serodiagnosis of Zika virus (ZIKV) infections by a novel NS1-based ELISA devoid of cross-reactivity with dengue virus antibodies: a multicohort study of assay performance, 2015 to 2016. Euro Surveill. 21:30426. doi: 10.2807/1560-7917.ES.2016.21.50.30426

PubMed Abstract | CrossRef Full Text | Google Scholar

St. George, K., Sohi, I. S., Dufort, E. M., Dean, A. B., White, J. L., Limberger, R., et al. (2017). Zika virus testing considerations: lessons learned from the first 80 real-time reverse transcription-PCR-positive cases diagnosed in New York State. J. Clin. Microbiol. 55, 535–544. doi: 10.1128/JCM.01232-16

PubMed Abstract | CrossRef Full Text | Google Scholar

Sumita, L. M., Rodrigues, J. P., Ferreira, N. E., Felix, A. C., Souza, N. C., Machado, C. M., et al. (2016). Detection of human anti-zika virus IgG by Elisa using an antigen from in vitro infected Vero cells: preliminary results. Rev. Inst. Med. Trop. Sao Paulo. 58:89. doi: 10.1590/s1678-9946201658089

PubMed Abstract | CrossRef Full Text | Google Scholar

Tan, S. K., Sahoo, M. K., Milligan, S., Taylor, N., and Pinsky, B. A. (2017). Stability of Zika virus in urine: specimen processing considerations and implications for the detection of RNA targets in urine. J. Virol. Methods. 248, 66–70. doi: 10.1016/j.jviromet.2017.04.018.

PubMed Abstract | CrossRef Full Text | Google Scholar

Tian, B., Qiu, Z., Ma, J., Zardán Gómez de la Torre, T., Johansson, C., Svedlindh, P., et al. (2016). Attomolar Zika virus oligonucleotide detection based on loop-mediated isothermal amplification and AC susceptometry. Biosens. Bioelectron. 86, 420–425. doi: 10.1016/j.bios.2016.06.085

PubMed Abstract | CrossRef Full Text | Google Scholar

Tognarelli, J., Ulloa, S., Villagra, E., Lagos, J., Aguayo, C., Fasce, R., et al. (2015). A report on the outbreak of Zika virus on Easter Island, South Pacific, 2014. Arch. Virol. 161, 665–668. doi: 10.1007/s00705-015-2695-5

PubMed Abstract | CrossRef Full Text | Google Scholar

Troncoso, A. (2016). Zika threatens to become a huge worldwide pandemic. Asian Pac. J. Trop. Biomed. 6, 520–527. doi: 10.1016/j.apjtb.2016.04.004

CrossRef Full Text | Google Scholar

Van den Hurk, R., and Evoy, S. (2015). A review of membrane-based biosensors for pathogen detection. Sensors. 15, 14045–14078. doi: 10.3390/s150614045

PubMed Abstract | CrossRef Full Text | Google Scholar

Waggoner, J. J., and Pinsky, B. A. (2016). Zika virus: diagnostics for an emerging pandemic threat. J. Clin. Microbiol. 54, 860–867. doi: 10.1128/JCM.00279-16

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, X., Yin, F., Bi, Y., Cheng, G., Li, J., Hou, L., et al. (2016). Rapid and sensitive detection of Zika virus by reverse transcription loop-mediated isothermal amplification. J. Virol. Methods. 238, 86–93. doi: 10.1016/j.jviromet.2016.10.010

PubMed Abstract | CrossRef Full Text | Google Scholar

Way, J. H., Bowen, E. T., and Platt, G. S. (1976). Comparative studies of some African arboviruses in cell culture and in mice. J. Gen. Virol. 30, 123–130. doi: 10.1099/0022-1317-30-1-123

PubMed Abstract | CrossRef Full Text | Google Scholar

Weaver, S. C., Costa, F., Garcia-Blanco, M. A., Ko, A. I., Ribeiro, G. S., Saade, G., et al. (2016). Zika virus: history, emergence, biology, and prospects for control. Antiviral Res. 130, 69–80. doi: 10.1016/j.antiviral.2016.03.010

PubMed Abstract | CrossRef Full Text | Google Scholar

Wilson, H. L., Tran, T., Druce, J., Dupont-Rouzeyrol, M., and Catton, M. (2017). Neutralization assay for zika and dengue viruses by use of real-time-PCR-based end point assessment. J. Clin. Microbiol. 55, 3104–3112. doi: 10.1128/JCM.00673-17

CrossRef Full Text | Google Scholar

Wong, S. J., Furuya, A., Zou, J., Xie, X., Dupuis, A. P., Kramer, L. D., et al. (2017). A multiplex microsphere immunoassay for Zika virus diagnosis. EBioMedicine 16, 136–140. doi: 10.1016/j.ebiom.2017.01.008

PubMed Abstract | CrossRef Full Text | Google Scholar

Xu, M. Y., Liu, S. Q., Deng, C. L., Zhang, Q. Y., and Zhang, B. (2016). Detection of Zika virus by SYBR green one-step real-time RT-PCR. J. Virol. Methods. 236, 93–97. doi: 10.1016/j.jviromet.2016.07.014

PubMed Abstract | CrossRef Full Text | Google Scholar

Yadav, S., Rawal, G., and Baxi, M. (2016). Zika virus: a pandemic in progress. J. Transl. Int. Med. 4, 42–45. doi: 10.1515/jtim-2016-0009

PubMed Abstract | CrossRef Full Text | Google Scholar

Yang, Y., Wong, G., Ye, B., Li, S., Li, S., Zheng, H., et al. (2017). Development of a reverse transcription quantitative polymerase chain reaction-based assay for broad coverage detection of African and Asian Zika virus lineages. Virol. Sin. 32, 199–206. doi: 10.1007/s12250-017-3958-y

PubMed Abstract | CrossRef Full Text | Google Scholar

Yaren, O., Alto, B. W., Gangodkar, P. V., Ranade, S. R., Patil, K. N., Bradley, K. M., et al. (2017). Point of sampling detection of Zika virus within a multiplexed kit capable of detecting dengue and chikungunya. BMC Infect. Dis. 17:293. doi: 10.1186/s12879-017-2382-0

PubMed Abstract | CrossRef Full Text | Google Scholar

Zaghloul, H., and El-shahat, M. (2014). Recombinase polymerase amplification as a promising tool in hepatitis C virus diagnosis. World J. Hepatol. 6, 916–922. doi: 10.4254/wjh.v6.i12.916

PubMed Abstract | CrossRef Full Text | Google Scholar

Zanluca, C., de Melo, V. C., Mosimann, A. L., Dos Santos, G. I., Dos Santos, C. N., and Luz, K. (2015). First report of autochthonous transmission of Zika virus in Brazil. Mem. Inst. Oswaldo Cruz 110, 569–572. doi: 10.1590/0074-02760150192

PubMed Abstract | CrossRef Full Text | Google Scholar

Zanluca, C., and Dos Santos, C. N. (2016). Zika virus an overview. Microbes Infect. 18, 295–301. doi: 10.1016/j.micinf.2016.03.003

PubMed Abstract | CrossRef Full Text | Google Scholar

Zare Mehrjardi, M., Keshavarz, E., Poretti, A., and Hazin, A. N. (2016). Neuroimaging findings of Zika virus infection: a review article. Jpn. J. Radiol. 34, 765–770. doi: 10.1007/s11604-016-0588-5

PubMed Abstract | CrossRef Full Text | Google Scholar

Zare Mehrjardi, M., Poretti, A., Huisman, T. A., Werner, H., Keshavarz, E., and Araujo Júnior, E. (2017). Neuroimaging findings of congenital Zika virus infection: a pictorial essay. Jpn. J. Radiol. 35, 89–94. doi: 10.1007/s11604-016-0609-4

PubMed Abstract | CrossRef Full Text | Google Scholar

Zaytseva, N. V., Montagna, R. A., Lee, E. M., and Baeumner, A. J. (2004). Multi-analyte single-membrane biosensor for the serotype-specific detection of Dengue virus. Anal. Bioanal. Chem. 380, 46–53. doi: 10.1007/s00216-004-2724-9

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhang, B., Pinsky, B. A., Ananta, J. S., Zhao, S., Arulkumar, S., Wan, H., et al. (2017). Diagnosis of Zika virus infection on a nanotechnology platform. Nat. Med. 23, 548–550. doi: 10.1038/nm.4302

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: Zika virus, Zika fever, diagnosis, surveillance, monitoring

Citation: Singh RK, Dhama K, Karthik K, Tiwari R, Khandia R, Munjal A, Iqbal HMN, Malik YS and Bueno-Marí R (2018) Advances in Diagnosis, Surveillance, and Monitoring of Zika Virus: An Update. Front. Microbiol. 8:2677. doi: 10.3389/fmicb.2017.02677

Received: 21 August 2017; Accepted: 22 December 2017;
Published: 19 January 2018.

Edited by:

Akio Adachi, Tokushima University, Japan

Reviewed by:

Muhammad Zubair Shabbir, University of Veterinary and Animal Sciences, Pakistan
Oscar Negrete, Sandia National Labs, United States

Copyright © 2018 Singh, Dhama, Karthik, Tiwari, Khandia, Munjal, Iqbal, Malik and Bueno-Marí. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Raj K. Singh, rks_virology@rediffmail.com
Kuldeep Dhama, kdhama@rediffmail.com

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.